Benha University

محمود محمد مجدى

مدرس

كلية الهندسة ببنها الهندسة الميكانيكية

Prototyping and Real-Time PID Control of a 5 DOF Hybrid Robotic-Assisted System for Neurosurgery

• 2025
العودة
معلومات البحث
المؤلفون A Sedky, A Marzuq, NA Mansour, A ElAssal, M Magdy
الكلمات المفتاحية Not Available
المجلة العلمية Not Available
الناشر Not Available
المجلد Not Available
العدد Not Available
الصفحات Not Available
publication.type International
رابط البحث Not Available
المواد المرفقة Not Available
الملخص
Neurosurgical interventions require high/exceptional/extreme accuracy due to the brain's intricate structure. Although robotic devices have improved accuracy in neurosurgery, attaining sub-millimeter precision and adaptability poses challenges. This study introduces a five-degree-of-freedom (DoF) hybrid robotic system that makes craniotomy and stereotactic procedures more reliable. The design has two Remote Centre of Motion (RCM) mechanisms: a 3-DoF main mechanism for aligning the tool and a 2-DoF compensating mechanism for moving the skull surface. Real-time joint control is guaranteed by a proportional-integral-derivative (PID) control method that was validated through MATLAB/Simulink hardware-in-the-loop (HIL) testing. The results of the experiments show that the maximum absolute error in trajectory tracking is 1.32 mm at the primary mechanism and 1° at the compensatory mechanism. The obtained results demonstrate proof of concept and confirm the design’s functional validity. The system's performance confirms its capability to improve neurosurgical precision while mitigating the spatial and stiffness constraints of current platforms