Mahmoud Mohamed Magdy
Lecturer
Prototyping and Real-Time PID Control of a 5 DOF Hybrid Robotic-Assisted System for Neurosurgery
• 2025
Publication Information
Authors
A Sedky, A Marzuq, NA Mansour, A ElAssal, M Magdy
Keywords
Not Available
Journal
Not Available
Publisher
Not Available
Volume
Not Available
Issue
Not Available
Pages
Not Available
publication.type
International
Paper Link
Not Available
Supplementary Materials
Not Available
Abstract
Neurosurgical interventions require high/exceptional/extreme accuracy due to the brain's intricate structure. Although robotic devices have improved accuracy in neurosurgery, attaining sub-millimeter precision and adaptability poses challenges. This study introduces a five-degree-of-freedom (DoF) hybrid robotic system that makes craniotomy and stereotactic procedures more reliable. The design has two Remote Centre of Motion (RCM) mechanisms: a 3-DoF main mechanism for aligning the tool and a 2-DoF compensating mechanism for moving the skull surface. Real-time joint control is guaranteed by a proportional-integral-derivative (PID) control method that was validated through MATLAB/Simulink hardware-in-the-loop (HIL) testing. The results of the experiments show that the maximum absolute error in trajectory tracking is 1.32 mm at the primary mechanism and 1° at the compensatory mechanism. The obtained results demonstrate proof of concept and confirm the design’s functional validity. The system's performance confirms its capability to improve neurosurgical precision while mitigating the spatial and stiffness constraints of current platforms